HAEMONETICS TEG 5000 PDF
The TEG® Hemostasis Analyzer System provides a more complete .. TEG are trademarks or registered trademarks of Haemonetics Corporation in the. Consult HAEMONETICS’s entire TEG ® catalogue on MedicalExpo. Page: 1/8. Consult HAEMONETICS’s entire TEG Hemostasis Analyzer System catalogue on MedicalExpo. Page: 1/8.
Author: | Daitilar Vudoran |
Country: | Mozambique |
Language: | English (Spanish) |
Genre: | Music |
Published (Last): | 11 January 2011 |
Pages: | 120 |
PDF File Size: | 4.45 Mb |
ePub File Size: | 19.2 Mb |
ISBN: | 674-1-86852-809-2 |
Downloads: | 4358 |
Price: | Free* [*Free Regsitration Required] |
Uploader: | Tojagami |
PN TEG User Manual
You can choose to view all tracings or filter the tracings by multiple criteria including patient name, test result ranges, site such as OR1, lab, etc. Certain common hemostasis conditions e.
Page 63 Undo Undo Chapter 4: Components of Hemostasis The ideal way to treat a bleeding or prothrombotic patient is to measure the net product of the multitude of the interacting factors and cellular elements and their interactions in the shortest time possible.
The program projects whether the MA will reach the lower limit of normal for certain sample types. Database operations your system is taking longer and longer to retrieve data. Press the index key A in Figure Similarly, in the case of tracing 5, a typical primary fibrinolysis pattern, where the R is slightly elongated and the MA is small and decreasing, fibrinolysis has to be treated before evaluating R, K, aand MA, unless these parameters show hypercoagulability, where R and K are small, and MA and a large.
Lastname, firstname [procedure] as in: Case Rx tab By inputting the values here with times, the case summary report trends will indicate these values on the timeline when displaying the test trend as well as in the numerical grid see below under Case summary report. This graphic is called the clot, and is displayed together with a normal representation of the clot to the left, as shown above in figure 4.
If no keyboard or mouse activity occurs for the time period haemonetlcs by your Site Administrator, the screen blacks out and displays the Operator Login screen.
If you click on the Patient ID field in the status bar, the Edit case screen is displayed and any patient information can be edited, as described below in the section named “Case Management. Note that these options are not available if you are logged in with Guest user group privileges.
Hemostasis p Antifibrinolytic drugs: Other clot formation parameters In addition to the major parameters just described, several others can aid in determining clot kinetics, strength, and stability: Local toolbar The toolbar at the top of the display panel local toolbar provides options for changing the size of tracings displayed Multi and Max as well tet for viewing filtered subsets of the database Filter, Active, SiteID, and Patient heamonetics displaying the graphic of the clot Data.
The corresponding tracings are displayed at the left. Despite the many components entering into hemostasis, two facts override all others: Sample detail screen accessed from Case samples tab.
When you click on the Print button, a dialog box is presented within which are options for printing page ranges. Creating a record If you want to create a record to contain manually entered data, click on Records in the Main menu, then New, then on Record. Report options You can also select Clot to print the clot graphic when reporting on a single sample. Therefore, K-time is a measure of the speed or clot kinetics to reach a certain level of clot strength.
Lift the carrier carefully to the pin with the lever still in the load position.
TEG Haemonetics U.S.A – HSC
For example, suppose that a patient sample produces a tracing as shown next: As the clot retracts or lyses, these bonds are broken and the transfer of cup motion is diminished. If you delete multiple samples, only the most recent deletion can be restored, and only before leaving the screen.
In the management of hemostasis, it is not typical to draw and analyze just one blood sample.
The utility of teh parameter is demonstrated by Szefner et al3 and Copeland et al4 in the monitoring of the hemostasis of patients undergoing total artificial heart or heart assist device implantation. This represents clot lysis. To turn off the reference tracing, click again on the Reference button in the toolbar. Print — page range option.
Educational Tools
CPT codes CPT codes for reimbursement purposes are listed at the bottom of the full report for a single sample they are not listed in multi-sample reports. They stabilize and significantly enhance the structure rigidity of the fibrin network. Page 1 Chapter 1: You must be in the TEG screen to start samples.
Case notes tab These notes are printed in the case summary report, as well as in the single sample report. However, by using the measurements and established normal ranges and indices, the patterns can be quantified as to the degree of abnormality, haeminetics described in Chapter 2. If you are running the Remote version, the QC prompt is disabled: Start Access, and follow these steps haeemonetics import your data Note that some of the prompts may vary, depending on which version of Access you are using: First, however, you should check to see whether it has been excluded from the listing.